• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE ALCON CLOSURE SYSTEM (ACS), NYLON; SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE ALCON CLOSURE SYSTEM (ACS), NYLON; SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE Back to Search Results
Catalog Number 8065710401
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A healthcare professional reported that, an ophthalmic suture was broken.Procedure details and patient impact have not been provided.Additional information has been requested, none received till date.
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information was provided in sections h.6 and h.11.A sample was not received at the manufacturing site for evaluation for the report of broken threads; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.Because a sample was not received and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, the root cause for customer complaint issue cannot be determined.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.Suture material samples are tested at incoming for tensile strength and diameter and visually inspected for any defects such as discoloration and frays.Sutures are also 100% inspected by trained operators for discoloration and frays and for proper attachment during the manufacturing process.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALCON CLOSURE SYSTEM (ACS), NYLON
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20532022
Report Number2523835-2024-02248
Device Sequence Number8247692
Product Code GAR
UDI-Device Identifier00380657104017
UDI-Public(01)00380657104017(17)270930(11)221007(10)153MF1(21)NA
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K930517
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065710401
Device Lot Number153MF1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/26/2024
Supplement Date Received by Manufacturer10/25/2024
12/23/2024
Initial Report FDA Received Date10/25/2024
Supplement Report FDA Received Date11/19/2024
01/17/2025
Date Device Manufactured10/07/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-