• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5 13.2
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Blurred Vision (2137)
Date of Event 09/21/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Manufacture narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.(b)(4).
 
Event or Problem Description
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.In a separate visit the lens was explanted and replaced for a replacement lens.The problem was resolved.Cause of the event was reported as unknown.
 
Additional Manufacturer Narrative
H3 - device evaluation: the lens was returned in liquid in a microcentrifuge vial.Visual inspection found haptic broken to the lens.Dimensional and functional inspection found the lens to be within specifications.Claim #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key20539175
Report Number2023826-2024-04932
Device Sequence Number8284374
Product Code QCB
UDI-Device Identifier00840311312269
UDI-Public(01)00840311312269(11)231109(17)261031
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberVTICM5 13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/09/2024
Supplement Date Received by Manufacturer01/28/2025
Initial Report FDA Received Date10/25/2024
Supplement Report FDA Received Date01/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: SFC-45 - LOT# UNK; FOAM TIP PLUNGER - LOT# UNK; INJECTOR MODEL: MSI-PF - LOT# UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient SexUnknown
-
-