| Model Number |
VTICM5 13.2 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
09/21/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Manufacture narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.(b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.In a separate visit the lens was explanted and replaced for a replacement lens.The problem was resolved.Cause of the event was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: the lens was returned in liquid in a microcentrifuge vial.Visual inspection found haptic broken to the lens.Dimensional and functional inspection found the lens to be within specifications.Claim #(b)(4).
|
| |
|