| Model Number |
VTICM5 13.2 (-14.0/1.0) |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
08/04/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6: type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced lens rotation not associated to a low vault.The lens remains implanted.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.Cause of the event was reported as unknown.
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced lens rotation and blurred vision.The lens was repositioned, and then exchanged for a different model lens of the same length.Additional notes reported, "lri performed at time of implantation." cause of event was reported as device and patient related factor.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: a1: patient identifier: kt, a2: date of birth: on (b)(6) 1988, a3a: sex: male, d4: model#: vticm5 13.2 (-14.0/1.0), catalog#: vticm5 13.2, serial#: (b)(6), primary unique device identifier (udi) #: (b)(4), g2: report source: health professional.Additional information: b1: adverse event.B2: required intervention to prevent permanent impairment/damage.B5: describe event or problem: a healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced lens rotation and blurred vision.The lens was repositioned, and then exchanged for a different model lens of the same length.Additional notes reported, "lri performed at time of implantation." cause of event was reported as device and patient related factor.D6b: if explanted, give date: on (b)(6) 2025.H1: type of reportable event: serious injury.H6: health effect - clinical code (e): 2137.H6: health effect - impact code (f): 4629.H6: medical device problem code (a): 2993.Claim#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation 3331 - device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within established process parameters and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim#: (b)(4).
|
| |
|
Search Alerts/Recalls
|