| Model Number |
EG-3670URK |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the balloon feeder return tube clogged.Based on the result, we concluded that it was caused due to the inadequate/insufficient reprocessing at the facility on the balloon feeder return tube.Based on the technical report "hr-rpt-0630(air/water & jet water channels)" and/or the risk analysis results, it was evaluated to submit mdr.
|
| |
|
Event or Problem Description
|
|
The time of event is unknown.There was no report of patient harm.Air/water tube clogged.
|
| |
|
Search Alerts/Recalls
|