Additional information provided in d.9, h.3, h.6, and h.11.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.Current tracking indicates no adverse trend for this lot for this event.The customer reported an aspiration failure occurred.The returned sample was visually inspected and functionally tested.Both irrigation and aspiration lures were intact.Flare shape was observed in the blue socket fitting at the aspiration tubing insertion end.The sample was tested using a calibrated console with the current software revision.The sample could be primed and tuned, and the service data could be retrieved from the console.The aspiration flow, irrigation flow, and irrigation met specification.The root cause of the customer's complaint could not be established since the returned sample met specifications.No contributing factors could be identified that could cause the reported complaint.After the investigation of this complaint, it has been determined that this sample met specifications.Therefore, no action will be taken at this time.Based on our current tracking, there are no adverse trends for this reported complaint.Quality assurance will continue to monitor customer complaints via the complaint review meetings and will take action for any future occurrences as is deemed necessary.The manufacturer internal reference number is: (b)(4).
|