• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW EVH ACCSESSORY PACK; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW EVH ACCSESSORY PACK; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VH-2004
Medical Device Problem Code Improper Flow or Infusion (2954)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
The hospital reported that during an endoscopic vein harvesting procedure, vv 6 accessory pack insufflator would not hold pressure inside the patient and the veins could not be visualized.Open vein harvesting was done.
 
Additional Manufacturer Narrative
Tw id#(b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Trackwise # (b)(4).Updated section: b4, d9, g3, g6, h2, h3, h6, h11.The lot # 3000363621 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW EVH ACCSESSORY PACK
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key20543474
Report Number2242352-2024-0001178
Device Sequence Number7873174
Product Code GEI
UDI-Device Identifier00607567700376
UDI-Public(01)00607567700376(10)3000363621(17)260101
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K041981
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2026
Device Model NumberVH-2004
Device Catalogue NumberVH-2004
Device Lot Number3000363621
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2024
Supplement Date Received by Manufacturer02/20/2025
Initial Report FDA Received Date10/28/2024
Supplement Report FDA Received Date02/21/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
Patient Weight105 KG
-
-