BIOMET UK LTD. CER OPTION TYPE 1 TPR SLEVE +6; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/02/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported patient underwent hip revision for unknown reasons; the surgeon explanted femoral head and acetabular components.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4): d10: item # 650-1057, cer bioloxd option hd 36mm, lot # 3062925.Item # xl-156233, arcomxl 36mm rlc lnr +5mm sz23, lot # 65781860.Item # unknow, unknown ringloc shell.Lot # unknown.Item # 00-7053-058-20, full bladeg 58mm, lot # unknown.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: g3; h2; h3; h6; h10 no product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Review of the complaint histories identified additional similar complaints for the reported items and no additional similar complaints for the reported part and lot combinations.Complaints are monitored per complaint trending process.Based on the available information, the reported event cannot be confirmed.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional information at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CER OPTION TYPE 1 TPR SLEVE +6
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key20544931
Report Number3002806535-2024-00408
Device Sequence Number12627645
Product Code LZO
UDI-Device Identifier00887868271502
UDI-Public(01)00887868271502(17)330309(10)3137180
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K192683
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number650-1068
Device Lot Number3137180
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2024
Supplement Date Received by Manufacturer11/06/2024
Initial Report FDA Received Date10/28/2024
Supplement Report FDA Received Date11/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-