|
This report is being supplemented to provide additional information based on the approved final investigation.New information has been added to the following fields: d8, d9, h3, h4, h6, h11.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue were identified.Based on the results of the investigation, the suggested event was caused since water or moisture invaded into the scope, thus they led to electronic components including the image sensor unit and the internal circuit board of the connector to give rise to defect.Important information ¿ please read before use.Precaution: caution.Turn the video system center on only when the endoscope connector is connected to the video system center.In particular, confirm that the video system center is off before connecting or disconnecting the endoscope connector.Failure to do so can result in equipment damage, including destruction of the image sensor.¦precautions for disappeared or frozen endoscopic image: warning follow the precautions given below.Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination.3.8 inspection of the endoscopic system.Inspection of the endoscopic image.Confirm that the wli, nbi, and rdi endoscopic images are normal.5.1 troubleshooting.If any irregularity is observed during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope and solve the problem as described in section 5.2,¿troubleshooting guide¿.If the problem still cannot be resolved, send the endoscope to olympus for repair as described in section 5.4, ¿returning the endoscope for repair¿.Also, should any irregularity be observed while using the endoscope, stop using it immediately and withdraw the endoscope from the patient as described section 5.3, ¿withdrawal of the endoscope with an irregularity¿.Olympus will continue to monitor field performance for this device.
|