| Catalog Number |
107140 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection of the photo showed a leak at the level of the anticoagulant line.Also observed was a drug syringe used during treatment that contained epoprostenol.Epoprostenol is not compatible with the polyethylene terephthalate glycol components of the prismaflex set.Per the set instructions for use, the drug should not be injected into the anticoagulation line as component damage can occur which may result in a leak.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the event was determined to be related to user error.A previous nonconformance was opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an external blood leak was observed from the filter of a prismaflex hf1000 set during hemodialysis therapy.The volume of blood loss was not known.There was no report of medical intervention associated with this event.No additional information is available.
|
| |
|