| Model Number |
3055601 |
| Medical Device Problem Codes |
Vibration (1674); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis of the device found that the handpiece was cosmetically not ok, nose cone assembly got corroded and motor, nose co ne and internal multiple spares found faulty.Section e: facility details are not available as on date of this report.Following is being performed and supplemental report will be submitted once additional information is received.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that pre-op, the skeeter drill had bur wobbling and was making noise.There was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported handpiece was a demo unit which is used by the training and education team of medtronic and there was no hospital involved.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|