• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR ARTOURA PLUS, SMOOTH, ULTRA HIGH PROFILE; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR ARTOURA PLUS, SMOOTH, ULTRA HIGH PROFILE; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number SDC135UH
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 09/28/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient who had a 700cc mentor artoura plus smooth ultra high-profile tissue expander placed experienced left sided deflation post procedure.Replacement with a permanent implant was performed on (b)(6) 2024.
 
Additional Manufacturer Narrative
The mentor failure analysis lab received the device for evaluation on november 14, 2024.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
The investigation of the returned device was completed by the failure analysis lab on february 26, 2025.Device evaluation summary: the product was returned to mentor for evaluation.Mentor then conducted a visual inspection, leak testing, and microscopic examination of the tissue expander.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the tissue expander.In addition, the tissue was received painted blue.Leak testing was performed, in accordance with mentor procedures, and a tear was observed at the junction between the shell and bladder at the eight o'clock position.Microscopic examination was performed and the cause of the deflation could not be identified.Based on the information currently available, a notification was sent to mentor's manufacturing team for further investigation.In conclusion, the collected process controls and data records for the deflated tissue expander meet specifications and the tear mentioned cannot be conclusively confirmed as a manufacturing defect.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: the silicone shell of the tissue expander may easily be cut by a scalpel or ruptured by excessive stress, manipulation with blunt instruments, or penetration by a needle.Subsequent deflation and/or rupture could result.All devices should be carefully inspected for structural integrity prior to and during implantation.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR ARTOURA PLUS, SMOOTH, ULTRA HIGH PROFILE
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key20546831
Report Number1645337-2024-12597
Device Sequence Number15407950
Product Code LCJ
UDI-Device Identifier10081317028236
UDI-Public(01)10081317028236(17)261012(10)9812559
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K161176
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSDC135UH
Device Lot Number9812559
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/08/2024
Supplement Date Received by Manufacturer11/14/2024
02/26/2025
Initial Report FDA Received Date10/28/2024
Supplement Report FDA Received Date12/06/2024
02/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient SexFemale
-
-