| Model Number |
VOCSN, ENGLISH |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device needed service.Upon evaluation of the device, ventec observed that the ventilator was unable to maintain peep (positive end expiratory pressure).There was no patient involvement associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the device was evaluated by ventec where the reported issue of the ventilator being unable to maintain peep (positive end expiratory pressure) was confirmed.Ventec replaced the exhalation control module (ecm) to resolve the reported issue.Proper device operation was then confirmed through functional and performance testing.The investigation determined that the cause of the reported issue was the ecm.
|
| |
|