Model Number CT96STSA
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 03/26/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported device is not available to be returned to the manufacturer for evaluation.An investigation into the root cause for the event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
On or about (b)(6) 2014, plaintiff underwent placement of an angiodynamics smartport product, reference number (b)(4), lot number 4728535.The device was implanted by dr.(b))6), md, at (b)(6) hospital, for the purpose of providing vascular access.On or about (b)(6) 2015, plaintiff was diagnosed with thrombus appearing round structures at the base of the right atrium and superior vena cava at (b)(6).Plaintiff's medical team determined that the smartport was the source of the thrombosis and that the defective port had to be removed.On or about (b)(6) 2015 plaintiff's defective port was removed by dr.(b)(6), md at (b)(6).
 
Additional Manufacturer Narrative
The customer's reported complaint description of the patient experienced thrombosis cannot be confirmed due to the patient centric nature of the serious adverse event (sae).No port or catheter tubing product was returned to angiodynamics for evaluation since there was no reported port device malfunction or performance issue, just patient sae of thrombosis.Thrombosis is cautioned in the device dfu as potential complication for an implanted port system.A device history record (dhr) review of the packaging/assembly/component lots revealed no quality related issues or manufacturing deficiencies at the time of manufacture.Labeling review: the directions for use which is supplied to the user with this port device, contains the following statements: warnings - absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.A blood return should be present prior to usage of device for any therapy or testing.- if the patient complains of pain, or if there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Precautions · for peripheral placement, irritation to the vein, resulting in postoperative thrombophlebitis, has been associated with guidewire and introducer insertion.· when using percutaneous introducers: to avoid blood vessel damage, do not allow the percutaneous introducer sheath to remain indwelling in the blood vessel without the internal support of a catheter or dilator.· if more than one drug is to be administered, between the individual drug applications, flush the system with 5 to 10 ml normal saline for injection to prevent drug interactions.· after any infusion, injection or bolus application, the system should be flushed with normal saline for injection or locked with a heparin solution per institutional protocol to prevent thrombotic occlusion of the catheter.Potential complications: use of an angiodynamics port system involves potential risks normally associated with the insertion or use of any implanted device or indwelling catheter including but not limited to: these complications are well documented in literature and should be considered when a venous access device is utilized.Air embolism clot formation fibrin sheath infection inflammation thromboembolism thrombophlebitis thrombosis a review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMART PORT CT
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key20547795
Report Number1317056-2024-00204
Device Sequence Number20790502
Product Code LJT
UDI-Device IdentifierH787CT96STSA0
UDI-PublicH787CT96STSA0
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K062414
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2015
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2017
Device Model NumberCT96STSA
Device Lot Number4728535
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2024
Supplement Date Received by Manufacturer11/18/2024
Initial Report FDA Received Date10/28/2024
Supplement Report FDA Received Date11/21/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/27/2014
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other; Hospitalization;
Patient SexMale
-
-