(b)(4).Date sent 10/28/2024.D4 batch #801c17.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned reload.Visual analysis of the returned sample determined that the sr55 reload was received with no apparent damage along with its opened packaging.The reload had the proximal 7 drivers up, the remaining drivers down with staples present, swing tab in the locked position and the knife not completely within the doghouse.The reload swing tab was reset in order to fire the cartridge.The returned reload was tested with a test device and fired without any difficulties.The staple line was complete, and the staples meet the staple form release criteria.The event reported was confirmed and it is related to improper use of the device.It should be noted that the reload is designed to lockout, as a safety feature, if any staples have been fired from the cartridge reload.In addition, failure to complete the stroke may result in an incomplete staple line.It is possible that the knife exposed noted on the reload is consistent with the firing knob was not returned completely prior to opening the device causing that the knife not returned completely to its home position.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 801c17, and no non-conformances were identified.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.
|