WRIGHT MEDICAL TECHNOLOGY INC INBONE EVERLAST SZ 2+ 12MM TOTAL ANKLE SYSTEM; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 22023212
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 10/01/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available for evaluation as it remains implanted in the patient therefore an evaluation of the device cannot be performed.If additional information becomes available, it will be provided on a supplemental report.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient will undergo a revision surgery due to severe lateral gutter impingement as well as the anterior positioning of the talar component.
 
Additional Manufacturer Narrative
Correction to d4 (gtin), d6a (date of implant), and h6(device code and conclusion code) the reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Upon further investigation of the ct scans, healthcare professionals opined that- the loosening and subsidence of the talus can be confirmed with the provided ct scan.The tibia, however, does not look loosened or otherwise altered.There is no clear radiolucence or cysts.The underlying cause is probably patient related due to poor bone substance.However; failure of total ankle replacement (tar) over time is a known complication.In case of a planned revision procedure a medical opinion aiming to determine the root cause of the failure is part of the internal investigation process.Sufficient (radiological and clinical) information must be provided to enable a meaningful clinical assessment and to identify possible causes of failure.In cases where a ct scan is the only available clinical source, this assessment is limited.No conclusive statement can be provided, because key clinical information (e.G.Clinical status, exact symptoms and range of motion of the patient, assessment of the treating physician) is missing.The root causes of radiographic findings such as radiolucent areas and bone cysts are often complex and multifactorial and cannot be determined scientifically without this decisive evidence.Based on the investigation the failure is most likely due to patient related issue i.E.Poor bone quality.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient will undergo a revision surgery due to severe lateral gutter impingement as well as the anterior positioning of the talar component.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INBONE EVERLAST SZ 2+ 12MM TOTAL ANKLE SYSTEM
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key20549588
Report Number3010667733-2024-00799
Device Sequence Number21077608
Product Code HSN
UDI-Device Identifier00889797074933
UDI-Public(01)00889797074933(11)201124(17)281122(10)1687155
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K193067
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number22023212
Device Lot Number1687155
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/01/2024
Supplement Date Received by Manufacturer12/26/2024
Initial Report FDA Received Date10/28/2024
Supplement Report FDA Received Date01/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/24/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age61 YR
Patient SexFemale
-
-