| Model Number |
Z800F |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/07/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device return requested.There are no previous complaints on this device.The complaint will be reopened and updated in the event the device involved or additional information becomes available.A follow-up mdr will be submitted in the event additional information becomes available.A capa has been opened in order to fully investigate and address the root cause of the reported event.Reference to complaint (b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, znyo medical received a report from the customer reporting that the pump infusing slower than programmed.The test results were at 6% the pump delivered 10.84ml.At -6% the pump delivered 10.01ml.No patient involved reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon reassessment, the reported flow rate failure mode has been determined to be non-reportable.Events involving a faster flow rate are not reportable when flow rate accuracy is above +5% but below +15% specification.The severity of this failure is 1 - inconvenience or temporary discomfort.Our evaluation concluded that the event did not pose a risk to the health of patients, users, or bystanders.Furthermore, the report did not allege a serious injury or death, nor would a recurrence of the reported issue be expected to cause or contribute to a serious injury or death.Reference to complaint #: (b)(4).
|
| |
|
Search Alerts/Recalls
|