DEPUY IRELAND - 3015516266 PINNACLE DM LINER TRIAL 50_43; HIP INSTRUMENTS : ACETABULAR TRIAL Back to Search Results
Catalog Number 221850043
Medical Device Problem Codes Material Deformation (2976); Naturally Worn (2988); Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that central sterile staff identified femoral head impactors with damaged tips, femoral canal probe with damaged flutes with the potential to produce metal burrs, and plastic trial components with deep grooves and damage.This did not happen in surgery.
 
Event or Problem Description
Additional information was received and states that: a.Were there any adverse consequence/s that affected the patient because of the reported event? no adverse consequences.B.Can you please clarify what you mean by "damaged"? was it worn, cracked, broken into pieces, bent, stripped, cross threaded, or any device interaction? worn, bent, and scratched.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information that was not available for the initial report is obtained, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> it was reported that central sterile staff identified femoral head impactors with damaged tips, femoral canal probe with damaged flutes with the potential to produce metal burrs, and plastic trial components with deep grooves and damage.Did not happen in surgery.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device did not found signs of stripped or deforming, however the pinnacle dm liner trial shows signs of several scratches around the surface.This type of damage is consistent with other tools and hard edges coming in contact with the device, properly handling and attention to the approved use of the device diminishes the risk of failure.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the pinnacle dm liner trial would have contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot
=
> the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative:.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE DM LINER TRIAL 50_43
Common Device Name
HIP INSTRUMENTS : ACETABULAR TRIAL
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key20550736
Report Number1818910-2024-22488
Device Sequence Number3114298
Product Code HWT
UDI-Device Identifier10603295520528
UDI-Public(01)10603295520528(10)AZ337789
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number221850043
Device Lot NumberAZ337789
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/23/2024
Supplement Date Received by Manufacturer10/29/2024
11/11/2024
12/03/2024
Initial Report FDA Received Date10/28/2024
Supplement Report FDA Received Date11/01/2024
11/13/2024
12/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/18/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-