| Model Number |
DSX900H11C |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/27/2024
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging a patient passed away due to the device.The device has not yet been returned to the manufacturer for evaluation.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging a patient passed away due to the device.The device has not yet been returned to the manufacturer for evaluation.The device was returned to the third-party service center for further evaluation.During the evaluation of the device at the third-party service centre, the device was visually inspected and found no evidence of foam degradation.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the third-party service centre.There was two error code found.And unit got scrapped.Box h updated.
|
| |
|