• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. BONE WAX 12 X 2.5GRAMS; WAX, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. BONE WAX 12 X 2.5GRAMS; WAX, BONE Back to Search Results
Catalog Number W31G
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported a patient underwent a breast surgery on (b)(6) 2024 and bone wax was used.The product does not adhere to the bone.Additional information has been requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional information: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.H6 component code: g07002 - device not returned.Additional information provided: the items from this batch are lighter/whiter than the more yellowish color it usually is.35 units in total - x9 items are opened and 26 still closed, but surgeons refusing to use products from this batch.Was surgery delayed due to the reported event? unknown.Action taken when event occurred? product from another company opened.Was procedure successfully completed? yes.Were fragments generated? no.If yes, were they removed easily without additional intervention? unknown.Patient status/ outcome / consequences: no.Was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: no.Is the patient part of a clinical study? unknown.Additional information has been requested however not received.Attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.Can you identify the lot number of the product that was used? please provide the source or name and title of the external person providing answers to follow-up (external person submitting answers to sales rep).Please provide the status of the device(s) as it has not been received for analysis.If the device has been shipped, please provide the shipment tracking details.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.Additional h6 component code: g07002 no device problem.Additional h3 investigation summary: the product was returned to ethicon for evaluation.Twenty-six representative unopened samples and four unused open samples were received for analysis.Product code w31g.As per the sampling plan, a visual inspection was performed on thirteen samples and the four open samples, the appearance and consistency of the bonewax bars were conforming the requirements.A functional test was performed on the bone wax, and it met the requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE WAX 12 X 2.5GRAMS
Common Device Name
WAX, BONE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-BRAZIL
rodovia presidente dutra
km 154
sao paulo 12240
BR   12240
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key20551131
Report Number2210968-2024-11176
Device Sequence Number15674969
Product Code MTJ
UDI-Device Identifier10705031056992
UDI-Public(01)10705031056992(17)290430(10)AV5545
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
PRE-AMEND
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberW31G
Device Lot NumberAV5545
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2024
Supplement Date Received by Manufacturer12/20/2024
Initial Report FDA Received Date10/28/2024
Supplement Report FDA Received Date01/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-