|
The complaint couldn't be confirmed, since the device was not returned for evaluation and no other evidences were provided.Upon further investigation of the ct scans by healthcare professionals the following was observed: there is some minor radiolucence adjacent to the talar component.There is, however, no clear loosening or migration of the talar component.Failure of total ankle replacement (tar) over time is a known complication.In case of a planned revision procedure a medical opinion aiming to determine the root cause of the failure is part of the internal investigation process.Sufficient (radiological and clinical) information must be provided to enable a meaningful clinical assessment and to identify possible causes of failure.In cases where a ct scan is the only available clinical source, this assessment is limited.No conclusive statement can be provided, because key clinical information (e.G.Clinical status, exact symptoms and range of motion of the patient, assessment of the treating physician) is missing.The root causes of radiographic findings such as radiolucent areas and bone cysts are often complex and multifactorial and cannot be determined scientifically without this decisive evidence.Due to these limitations, we are unable in such cases to provide statements about the patient, the procedure and/or the device in relation to cause of the failure.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|