| Model Number |
Z800WF |
| Medical Device Problem Codes |
Excess Flow or Over-Infusion (1311); Calibration Problem (2890); Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There are no previous complaints on the device.This pump underwent routine maintenance testing where it was detected the pump's performance fell outside the acceptable range for the flowrate %.Consequently, it necessitated an adjustment to the pump's flowrate %.It was confirmed the recalibration from 7 to 1 addressed the issue successfully.A capa has been opened in order to fully investigate and address the root cause of the reported event.Reference to complaint # (b)(4).
|
| |
|
Event or Problem Description
|
|
On 10/15/2024, znyo medical received a report that during the preventive maintenance (pm), the device was reported to have a flow rate offset issue.Flow rate deviation was too high (5.8).No report patient was involved.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplement serves as a correction as part of our retrospective 2024 complaint remediation.Upon reassessment, flow rate failures +5% to + 15% or -5% to -15% would not lead to the harm of a patient.The reported flow rate failure mode has been determined to be non-reportable.The severity of this failure is 1 - inconvenience or temporary discomfort.Reportability assessment: our evaluation concluded that this complaint reported did not pose risk to health to patient, users or bystanders, did not allege a serious injury or death and would not cause or contribute to a serious injury or death if the reported issue recurred.This event is not reportable.Reference to complaint # (b)(4).
|
| |
|