| Catalog Number |
363419 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: detailed inquiry description: while priming with blood product (plts) pump indicated "dowstream pressure".Upon assessing tubing clamps and cvad clamp, the blood tubing ruptured at proximal y port, while still connected to picc.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).No sample and one photograph was provided for evaluation.Upon visual inspection of the photograph, it was determined that a proper evaluation could not be performed from the evidence in the photograph provided.Based on the data from the investigation, the reported defect was unable to be confirmed.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|