• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ANTHOLOGY HO POR PL HA SZ 11; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ANTHOLOGY HO POR PL HA SZ 11; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE Back to Search Results
Catalog Number 71357111
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Periprosthetic fracture (4902)
Date of Event 10/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a thr surgery on (b)(6) 2016, the patient experienced a periprosthetic hip fracture.This adverse event was resolved by a revision surgery on (b)(6) 2024, where the anthology stem, femoral head, and liner of the r3 cup were removed.No further information is available at the moment.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H6: e2127 periprosthetic fracture.H3,h6: this complaint was opened by smith+nephew to document a patient complication reported that includes reference to the use of a smith+nephew product without evidence about a specific product problem.The reported complication relates to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
B5, d10, h3, h6: the devices were not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, it was communicated that the requested clinical documentation was not available.The general health of the patient and further details of the event are unknown.A definitive clinical root cause of the reported event cannot be concluded based on the limited information provided.Patient impact beyond the reported revision cannot be determined.A review of the production orders did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part numbers over the past 12 months and for the batch numbers based on historical data of the devices did not reveal similar events for the listed devices.A review of the instructions for use documents for total hip systems revealed that congenital deformity, osteoporosis, bone defects, trauma, strenuous activity, patient non-compliance, etc.Can increase risk of femoral or pelvic fractures.A review of the risk management files revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to these products and event.At this time, we have no evidence to conclude that the products failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include patient bone quality and/or or traumatic injury.Based on this investigation, the need for corrective action is not indicated.Should the devices or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Event or Problem Description
It was reported that, after a thr surgery on (b)(6) 2016, the patient experienced a periprosthetic hip fracture.This adverse event was resolved by a revision surgery on (b)(6) 2024, where the anthology stem, femoral head, and liner of the r3 cup were removed.No further information is available at the moment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY HO POR PL HA SZ 11
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key20556026
Report Number1020279-2024-02217
Device Sequence Number16934726
Product Code MRA
UDI-Device Identifier03596010559661
UDI-Public3596010559661
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/12/2025
Device Catalogue Number71357111
Device Lot Number15KM11275
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/01/2024
Supplement Date Received by Manufacturer11/25/2024
Initial Report FDA Received Date10/29/2024
Supplement Report FDA Received Date11/27/2024
Date Device Manufactured10/15/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OXINIUM FEM HD 12/14 36 MM +0.; PN: 71332760 / LOT: 16AM07288.; PN: 71343600 / LOT: 16AM07288.; R3 0 DEG XLPE ACET LNR 36MM X 60MM.
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient Age86 YR
Patient SexMale
-
-