• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #6; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #6; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6021-0637
Medical Device Problem Codes Material Erosion (1214); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Pain (1994); Ambulation or Postural Difficulties (2544)
Date of Event 10/01/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reported event: an event regarding disassociation and wear involving an accolade stem was reported.The event was confirmed via photographs provided.Method & results: product evaluation and results: visual inspection: the device was not returned; however, photographs were provided for review.These images clearly show the device in a disassociated and used state, with significant wear evident on the trunnion, and typical wear patterns from implantation and explantation.Based on this evidence, the event can be confirmed clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: no other similar events were reported for the lot.Conclusions: it was reported that the patient was revised due to disassociation and wear.The device was not returned; however, photographs were provided for review.These images clearly show the device in a disassociated and used state, with significant wear evident on the trunnion, and typical wear patterns from implantation and explanation.Based on this evidence, the event can be confirmed.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Patient had a right hip revision due to pain, noise in hip and the head disassociating from the stem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE PLUS TMZF HIP STEM #6
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
kelsey williams
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key20556590
Report Number0002249697-2024-01472
Device Sequence Number8979986
Product Code LPH
UDI-Device Identifier04546540510716
UDI-Public04546540510716
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2021
Device Catalogue Number6021-0637
Device Lot Number36208104
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/01/2024
Initial Report FDA Received Date10/29/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/21/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age84 YR
Patient SexMale
-
-