ASCENSION ORTHOPEDICS, INC. 38MM REVERSE LINER +0 S; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Catalog Number LNR-0990-200-380S
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 09/30/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a tsa surgery had been performed on (b)(6) 2024, the patient experienced a shoulder dislocation around on (b)(6) 2024.This adverse event was treated with a revision surgery on (b)(6) 2024, in which the 38mm reverse liner +0 s and the 38mm glenosphere concentric were exchanged for a 42mm glenosphere with the corresponding poly liner.Upon extraction no dissociation or implant malfunction was noted.Current health status of patient is unknown.
 
Additional Manufacturer Narrative
This complaint was opened by smith+nephew to document a patient complication reported that includes reference to the use of a smith+nephew product without evidence about a specific product problem.The reported complication relates to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
38MM REVERSE LINER +0 S
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ASCENSION ORTHOPEDICS, INC.
11101 metric blvd
austin TX 78758
Manufacturer (Section G)
ASCENSION ORTHOPEDICS, INC.
11101 metric blvd
austin TX 78758
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key20556670
Report Number3002788818-2024-00111
Device Sequence Number14265471
Product Code PHX
UDI-Device Identifier00885556810170
UDI-Public885556810170
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K220847
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date07/01/2026
Device Catalogue NumberLNR-0990-200-380S
Device Lot Number700007165
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2024
Initial Report FDA Received Date10/29/2024
Date Device Manufactured07/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-