Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that the surgeon monitor could not display anything intermittently.Tested with another cable, and the issue was not solved.It was then tested with another monitor, which solved the issue.Software was also upgraded.There was no patient involvement.
|
|
Section d references the main component of the system.Other medical products in use during the event include: brand name ; product id 9733623 (lot: -); h3, h6: a medtronic representative went to the site to test the equipment.Testing confirmed that the monitor would not display anyt hing.The monitor was replaced to resolve this issue.Codes b01, c07 and d02 are applicable to this analysis. medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|