A batch review of the finished good lot and components confirmed there were no quality issues noted throughout the incoming inspection, manufacturing, sterilization, in-process, or final inspection.The needle component is supplied by ethicon.To avoid damaging needle points and swage areas, grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the swaged end to the point.Reshaping needles may cause them to lose strength and be less resistant to bending and breaking.Users should exercise caution when handling surgical needles to avoid inadvertent needle sticks as indicated in the precautions section of the instruction for use.The actual device was not available for review.Without receiving sterile devices from the same finished the ethicon third party evaluation of the needle component or receiving details regarding the preoperative preparation of the device, tools utilized to grasp the needle component, the size trocar used compared to the size needle on the device, procedure performed or the surgeon¿s technique, a definitive root cause cannot be determined at this time.
|