|
SMITH & NEPHEW, INC. LEGION AP BLOCK W/CRUCIATE SPIKES ¿ SZ 6; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
E0015293-5 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a tka surgery, the fixation post of a legion ap block w/cruciate spikes sz 6 broke off.All pieces were retrieved, but it is unknown how.The procedure was resumed, without any delay, with the same device, as all cuts were done prior to post breaking off.No injury was reported as a consequence of this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
H3,h6: the associated device was returned and evaluated.The visual inspection revealed that one of the spikes is fractured off the device.The fractured piece was returned.The device has scratches and gouges and is worn from use.This inspection was conducted by the naked eye.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.There is no complaint history for this part, as this is a custom-made device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.D4: serial #, lot #.
|
| |
|
Search Alerts/Recalls
|
|
|