| Model Number |
WA2UR23A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported, the uteteroscope's tip had four inches break off.The issue occurred during reprocessing.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Additional information: h3, h6, h11.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the reported event can be attributed to improper handling and application of excessive force, impact to the device like fall, shock or similar stress.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|