| Model Number |
V273 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This cardiac resynchronization therapy pacemaker (crt-p) was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.The device could not be interrogated and was exhibiting no pacing pulses.The device case was removed to facilitate inspection and testing of the internal components.The battery was removed and replaced with an external power source.Testing confirmed a normal current drain.The battery was forwarded for detailed analysis and while the battery was confirmed to be depleted, the root cause was unable to be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) was explanted and returned without an allegation.
|
| |
|