• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO GOBED II MED/SURG BED; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO GOBED II MED/SURG BED; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Catalog Number FL28C
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Codes Fall (1848); Laceration(s) (1946); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.4 devices were functionally/visually inspected in the field.The devices were repaired and returned to use.There was no remedial action taken.This device is not labeled for single use.
 
Event or Problem Description
This report summarizes 4 malfunction events, where it was reported the devices experienced reduced/inadequate brake force.There was 1 event with patient involvement; the patient experienced a fall and laceration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GOBED II MED/SURG BED
Common Device Name
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
melissa simon
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key20560253
Report Number0001831750-2024-01358
Device Sequence Number3088534
Product Code FNL
UDI-Device Identifier07613327169355
UDI-Public07613327169355
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized4
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberFL28C
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/01/2024
Initial Report FDA Received Date10/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-