• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INC. NUVASIVE MODULUS ALIF SYSTEM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INC. NUVASIVE MODULUS ALIF SYSTEM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number D1300096
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/18/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No device was returned for evaluation and no photographs of the device were provided for review.No operative notes were provided for review of usage/technique.Based on the information available, the complaint was unable to be confirmed and the cause of the reported event was unable to be determined conclusively; however, excessive use of force, off-axis force, technique, or material wear/fatigue may have caused or contributed to the reported event.Labeling review: ".Warnings, cautions and precautions: the implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient." ".The instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices.The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury." ".Care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." ".The physical characteristics required for many instruments do not permit them to be manufactured from implantable materials.If any broken fragments of instruments remain in the body of a patient, they could cause allergic reactions or infections.If an instrument breaks in surgery and fragments go into the patient, these pieces should be removed prior to closure and should not be implanted.".
 
Event or Problem Description
It was reported that during a procedure, the tip of the driver fractured off within the screw.The tip was left in-situ, embedded in the screw.No further patient impact was reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE MODULUS ALIF SYSTEM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
NUVASIVE, INC.
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INC.
7475 lusk boulevard
san diego CA 92121
Manufacturer Contact
macy stout
7475 lusk boulevard
san diego, CA 92121
MDR Report Key20562102
Report Number2031966-2024-00283
Device Sequence Number3200717
Product Code LXH
UDI-Device Identifier00195377060143
UDI-Public195377060143
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberD1300096
Device Lot NumberIN2756
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2024
Initial Report FDA Received Date10/29/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/10/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-