• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. SIMPLYGO; GENERATOR, OXYGEN, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. SIMPLYGO; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number 1068987
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/19/2024
Type of Reportable Event Death
Event or Problem Description
The manufacturer was contacted in reference to a customer requesting information on returning a simply go device due to a patient passing.There is no allegation of device malfunction.There is no allegation that the device contributed to the patient death.No other information is available at this time.This report is being submitted for an allegation of a patient death only due to insufficient information related to the device.The device has not been returned to the manufacturer for investigation.The manufacturer investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Additional Manufacturer Narrative
The manufacturer previously reported in reference to a customer requesting information on returning a simply go device due to a patient passing.There is no allegation of device malfunction.There is no allegation that the device contributed to the patient death.No other information is available at this time.This report is being submitted for an allegation of a patient death only due to insufficient information related to the device.The device was returned to a third-party service center.During the evaluation of the device, the service technician noted that the pca does not detect o2% level.The pca was replaced and all functional and quality checks passed.The device was scrapped by the service center after the evaluation was completed at the customer¿s request.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLYGO
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key20563622
Report Number2518422-2024-102889
Device Sequence Number7532089
Product Code CAW
UDI-Device Identifier00606959005723
UDI-Public(01)00606959005723
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111885
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number1068987
Device Catalogue Number1068987
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2024
Supplement Date Received by Manufacturer04/02/2025
Initial Report FDA Received Date10/29/2024
Supplement Report FDA Received Date08/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/17/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexUnknown
-
-