|
Updated h3 and h4 this report is being supplemented to provide additional information based on the approved final investigation.Based on the results of the investigation, as the device was not returned, the reported malfunction could not be reproduced.The likely cause could be that the water accumulated in the reprocessing basin due to leakage from the water supply valve.The possible causes of the leakage are as follows: the water supply hose was loosely attached due to the breakage of the water supply valve.However, a definitive root cause could not be determined.The event can be detected and prevented in accordance with the following instructions for use (ifu).[installation manual] chapter 5: installation of equipment 5.2 connection of the water supply hose.11.Open the water faucet slowly while checking to make sure that water does not leak from the hose connection areas (around the water supply plug, the faucet side plug, the equipment side plug, and the connection ring).If water leaks, close the water faucet and reconnect the water supply hose 1 and the water supply hose 2.[operation manual] chapter 8: troubleshooting and repair.If any irregularity is detected during an inspection or if the device is clearly malfunctioning, do not use it.Contact olympus for repair.Some problems that appear to be malfunctions may be correctable by referring to section 8.1, ¿troubleshooting guide,¿ on page 213.If the problem cannot be resolved by the described remedial action, do not use the device and contact olympus.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Olympus will continue to monitor the field performance of this device.
|