| Model Number |
3189 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Inadequate Pain Relief (2388); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event estimated.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's system diagnostics recorded high impedances on the supra-orbital/occipital leads.As a result, the patient was experiencing ineffective therapy.It was also reported the patient had a fall in (b)(6) 2024.Surgical intervention may take place at a future date to address the issue.
|
| |
|
Event or Problem Description
|
|
Additional information was received stating programming was able to provide effective therapy, no intervention took place.Therefore, decision is not reportable.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event estimated.Correction - section b5 - programming was able to provide effective therapy.Decision is no longer reportable.The results of the investigation are inconclusive since the devices were not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.
|
| |
|