| Catalog Number |
0240060100 |
| Medical Device Problem Codes |
No Display/Image (1183); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: flickering.Probable root cause: ¿ dvi / s-video/compositecables and connectors, ¿ input video signals/ recording formats from video source, ¿ sk28 system design (sk28-io board, sk28-m board), ¿ sk28 firmware/ software, ¿ sdc3 application software.¿ gravity workflow software application.¿ software: sdc3 ipad app.¿ use error and ¿ manufacturing/ service nonconformity.The reported failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Search Alerts/Recalls
|