MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BALLOON BIPOLAR TEMPORARY PACING ELECTRODE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BALLOON BIPOLAR TEMPORARY PACING ELECTRODE CATHETER Back to Search Results
Catalog Number 006172P
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/10/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the faulty temporary pacing electrode.Per follow up information received via ibc on 28oct2024, stated that the 2 yellow and black bits in the temporary pacing electrode should not be loose.
 
Event or Problem Description
It was reported that the faulty temporary pacing electrode.Per follow up information received via ibc on 28oct2024, stated that the 2 yellow and black bits in the temporary pacing electrode should not be loose.
 
Additional Manufacturer Narrative
The reported event was confirmed cause unknown.No actions can be taken at this time since a root cause was not identified.Visual evaluation of the returned one-photo sample noted one opened (without original packaging), temporary pacing electrode.Visual inspection of the one-photo sample noted that the cover of the electrode and the sleeve for the wire did not meet, there was a gap in the electrode.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: indications for use bard temporary pacing catheters are designed to transmit an electrical signal from an external pulse generator to the heart or from the heart to a monitoring device.When an internal lumen is present (other than the one used for balloon inflation), it may be used for fluid infusion, pressure monitoring, or blood sampling.Contraindications none.Warnings general warnings these warnings apply to all bard temporary pacing electrode catheters.Inappropriate electrical connections, e.G.Into a wall socket, may pose serious risk of adverse health consequences or death.Please ensure that the catheter is connected as recommended for pacing or measuring intracardiac electrograms.This device should be used only by or under the supervision of physicians trained in the techniques of transvenous intracardiac studies and temporary pacing.This device is for one time use only.Reuse or resterilization can impair the structural integrity and/or performance of the catheter.Adverse patient reactions can also result from reuse.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.The risks of using temporary pacing catheters include those risks related to heart catheterization, such as thromboembolism, perforation, tamponade, and infection.The induction of an unintended arrhythmia is a known complication.Instructions for use inspection instructions 1.Inspect the sterile package carefully for damage during transit or storage.Do not use the catheter if the package is damaged.2.Visually inspect the catheter, under sterile conditions, for kinks in the catheter shaft, integrity of the connector, condition of electrodes, and any other damage.3.In case of catheters with a balloon, under sterile conditions, remove the protective sheath and inflate the balloon with 1.5 ml of air or carbon dioxide.Use the inflation syringe included in the package.Completely deflate the balloon after the test.Insertion instructions using a needle cannula 1.Open the package and place the contents on a sterile field.2.Prep the skin at the site of insertion and inject a local anesthetic.3.Remove the protective guard from the needle cannula.4.Enter the vein with the needle cannula.Simultaneous aspiration into a syringe will help confirm vessel entry.5.Remove the syringe and the needle.6.If using an open-lumen catheter, flush the catheter with a heparinized solution.Remove any stylette prior to insertion.7.Using the aid of fluoroscopy or an ecg monitor, advance the catheter through the cannula to the desired position.If using a balloon catheter, inflate the balloon when the catheter is in the right atrium.Please note that the balloon can be inflated or deflated only when the stopcock is parallel to the catheter shaft.Do not pull the catheter back through the cannula as it may cause damage to the catheter.8.If using a balloon catheter, deflate the balloon after the catheter has reached the desired location.9.Test the pacing characteristics for optimal pacing.10.Pull the cannula back and secure it to the proximal end of the catheter.11.Secure the electrode catheter in place at the insertion site.Nsertion instructions using a percutaneous introducer sheath follow the instructions, warnings, and precautions of the introducer manufacturer.If using a balloon temporary pacing catheter, use half or one french size larger introducer, unless otherwise recommended by the introducer manufacturer.Electrical connections for measuring intracardiac ecgs 1.Insert adapter pin into standard 2 mm (0.080) pin receptacle of equipment (see figure 1).If equipment contains a locking mechanism such as a collet or thumbscrew, tighten down onto adapter.Leave affixed to equipment.Warning: inappropriate electrical connections, e.G.Into a wall socket, may pose serious risk of adverse health consequences or death.2.Thread leads of catheter through the adapter eyelet.Connect the negative jack (marked distal) to the v-lead of the ecg, and the positive jack (unmarked) to the positive terminal of the external pulse generator.Electrical connections for pacing 1.Insert adapter pin into standard 2 mm (0.080) pin receptacle of equipment (see figure 1).If equipment contains a locking mechanism such as a collet or thumbscrew, tighten down onto adapter.Leave affixed to equipment.Warning: inappropriate electrical connections, e.G.Into a wall socket, may pose serious risk of adverse health consequences or death.2.Thread leads of catheter through the adapter eyelet.Connect the negative jack (marked distal) to the negative terminal of the external pulse generator, and the positive jack (unmarked) to the positive terminal of the pulse generator.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLOON BIPOLAR TEMPORARY PACING ELECTRODE CATHETER
Common Device Name
TEMPORARY PACING ELECTRODE CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key20567155
Report Number1018233-2024-06883
Device Sequence Number7481533
Product Code LDF
UDI-Device Identifier00801741011191
UDI-Public(01)00801741011191(11)240501(17)260331(10)GFJQ2094
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K800298
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Catalogue Number006172P
Device Lot NumberGFJQ2094
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/09/2024
Supplement Date Received by Manufacturer03/12/2025
Initial Report FDA Received Date10/30/2024
Supplement Report FDA Received Date03/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-