DEPUY SYNTHES PRODUCTS LLC 6.5CM PEDIATRIC CRANIOTOME; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE
|
Back to Search Results |
|
| Catalog Number |
CRANI-P |
| Medical Device Problem Codes |
Break (1069); Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by japan that during service and evaluation, it was determined that the neuro-tip of the craniotome device bent/damaged.It was further observed that the device had cosmetic damage, worn bearing and vibration.It was further determined that the device failed pretest for visual assessment, color band visual assessment, cutter insertion and vibration.It was noted in the service order that the bearing was worn.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in (b)(6) 2024.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the craniotome neuro-tip being bent/damaged, identified during service and evaluation was confirmed.The assignable root cause was determined to be due to component failure from normal wear.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Correction: d9: date device returned to manufacturer: the device returned date was incorrect the initial medwatch report.The date has been updated accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|