• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, 1688 AIM 4K CAMERA HEAD WITH INTEGRATED COUPLER; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, 1688 AIM 4K CAMERA HEAD WITH INTEGRATED COUPLER; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 1688610122
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/04/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that there was blurry image.
 
Event or Problem Description
It was reported that there was blurry image.
 
Additional Manufacturer Narrative
Alleged failure: blurry image confirmed failure: no problem found probable root cause: cables hdmi cables connectors digital board power supply filter / fuse ac inlet board main board transition board intermittence between transition board and ch connector software camera head (sensor, sensor cable) coupler (dust on optics, scratched optics, broken / shifted lenses, focusing mechanism, endoclamp, zoom ring) problem toggling light source connected monitors leaking poor grounding (e.G camera head connection from cable to transition board.) no/incorrect communication with light source soaking cap disconnection during sterilization electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge cybersecurity attack pendulum ch inertial measurement unit (imu) incident light from surgical scene is reflected by coupler/ch window use error the product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 1688 AIM 4K CAMERA HEAD WITH INTEGRATED COUPLER
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key20569700
Report Number0002936485-2024-00734
Device Sequence Number20830085
Product Code GCJ
UDI-Device Identifier07613327418996
UDI-Public07613327418996
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K182160
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1688610122
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2024
Supplement Date Received by Manufacturer10/04/2024
Initial Report FDA Received Date10/30/2024
Supplement Report FDA Received Date01/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-