| Model Number |
G6623511 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/02/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6: neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient who underwent c5/6 anterior cervical discectomy and fusion for cervical spondylotic myelopathy.It was reportedthat the caudal screw was cut out after surgery.Additional surgery planned to remove the screw.There were no patient symptoms reported.There were no further complications reported regarding the event.Additional information received that the cage has moved from the right place to wrong place.Revision surgery was performed on october 3.
|
| |
|
Additional Manufacturer Narrative
|
|
Radiographic image review comments - 1.Sagittal ct reconstruction shows a screw in the anterior prevertebral space not engaged in the bone.2.Sagittal ct view shows superior screw of stand alone interbody connected device partially engaged in bone; whole graft appears pos itioned anteriorly.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|