• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. 12MM XCEL STABILITY SLEEVE; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. 12MM XCEL STABILITY SLEEVE; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Catalog Number CB12LT
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 10/30/2024.D4: batch # unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unknown procedure, the tip of the sleeve was bent.Changed to a new device.No patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 12/4/2024 d4: batch # unk investigation summary the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the cb12lt device was received with the tip of the sleeve damaged melted.In addition, the packaging opened was returned along with the instrument.The event reported was confirmed and it is related to improper use of the device.In order to prevent this kind of damage please avoid the contact with laser, electrosurgical or ultrasonic devices.A manufacturing record evaluation was performed for the finished device lot 772c47 number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12MM XCEL STABILITY SLEEVE
Common Device Name
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key20570820
Report Number3005075853-2024-08248
Device Sequence Number8095254
Product Code GCJ
UDI-Device Identifier10705036001331
UDI-Public(01)10705036001331(17)281130(10)772C47
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K032676
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCB12LT
Device Lot Number772C47
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/16/2024
Supplement Date Received by Manufacturer11/20/2024
Initial Report FDA Received Date10/30/2024
Supplement Report FDA Received Date12/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-