D10 concomitant product: cap, cap cap for use with fenestrated i x10 (lot # unknown) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that four samples were received for evaluation.All the samples were sealed inside their pouches.The lot numbers of the tubes were three samples of lot 23b1186jzx, and one sample of lot 24e1134jzx.It was observed that none of the tubes have the connector attached, although the connectors were present inside the pouches.The insertion mark of the connector in the proximal end was out of specification.A dimensional test was performed, and the inner and outer diameter of the tubes and the connectors were measured.It was reported that the cap of the endotracheal tube kept popping off when on a patient.The reported issue was confirmed.The most likely cause was determined to be manufacturing related.The instructions included with this device provide the following guidance: the 15 mm connector is seated so that it can be removed with effort if cutting of the tube is desired.Follow the instructions if cutting is considered.Do not use lubricating jelly to ease connector reinsertion, as it may contribute to accidental disconnection.If shortening of the endotracheal tube is needed, use medical judgment to determine the appropriate tube length for each individual patient.Before intubating the patient, cut the tube, reinsert the 15 mm connector, and verify the connector is firmly seated in the tracheal tube.If the device is damaged or the integrity of the sterile barrier has been compromised, do not use the product and contact your covidien representative for further information.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|