NUVASIVE INC. SIMPLIFY CERVICAL ARTIFICIAL DISC; PROSTHESIS INTERVERTEBRAL DISC Back to Search Results
Model Number MD-5L
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cyst(s) (1800); Neck Pain (2433); Metal Related Pathology (4530)
Date of Event 10/08/2024
Type of Reportable Event Serious Injury
Event or Problem Description
On approximately on (b)(6) 2024, a patient underwent a cervical total disc replacement (ctdr) procedure at c4/5.On (b)(6) 2024, a revision procedure was completed during and during the explant of the ctdr there were clear signs of metallosis which caused a cyst of fluid build up in the patient vertebral body causing severe pain and resulted in a fusion procedure.
 
Additional Manufacturer Narrative
The device was explanted and has been sent directly to a third party lab for evaluation however, radiographs provided confirm what appear to be cysts in the c7 vertebral body.No lot code information was provided so a manufacturing review could not be completed.He patient bone quality is unknown.Surgical notes stated blood work was negative for infection.The patient postoperative physical activity is unknown.With the current information provided a definitive root cause could not be determined although patient selection, implant selection and patient related factors may be a cause or contributors to the event.Once the external explant evaluation is complete an additional report will be filed.Labeling review: "contraindications simplify® cervical artificial disc should not be implanted in patients with the following conditions: an active systemic infection or an infection at the operative site.Osteoporosis or osteopenia defined as dexa bone mineral density t-score less than -1.5.Known allergy to the implant materials (peek, ceramic, titanium)." "warnings simplify® cervical artificial disc should only be used by surgeons who are experienced with anterior cervical spinal procedures and have undergone hands-on training in the use of this device.Only surgeons who are familiar with simplify® cervical artificial disc components, instruments, procedure, clinical applications, biomechanics, adverse events (aes), and risks associated with simplify® cervical artificial disc should use this device.A lack of adequate experience and/or training may lead to a higher incidence of aes, including neurological complications.Correct selection of the appropriate implant size and correct placement of simplify® cervical artificial disc are essential to ensure proper performance and functioning of the device.Information regarding implant size selection and correct implant placement is provided in the simplify® cervical artificial disc surgical technique guide." "precautions the safety and effectiveness of the simplify® cervical artificial disc has not been established in patients with the following conditions: previous surgery at the level to be treated.More than one neck surgery via anterior approach.Axial neck pain as the solitary symptom.History of an endocrine or metabolic disorder (e.G., paget¿s disease) known to affect bone and mineral metabolism.Current or extended use (>6 months) of any drug that may interfere with bony/soft tissue healing.Insulin-dependent diabetes.Renal disease.Auto-immune diseases." "preoperative: in order to minimize the risk of atraumatic periprosthetic vertebral fractures, surgeons must consider all co-morbidities, past and present medications, previous treatments, etc.Upon reviewing all relevant information,the surgeon must determine whether a bone density (dexa) scan is prudent.If dexa is performed, the patient should not receive the device if the dexa bone mineral density t-score is <-1.5, as the patient may be osteoporotic or osteopenic.Patient selection is extremely important.In selecting patients for a total disc replacement, the following factors can be of extreme importance to the success of the procedure: the patient¿s occupation or activity level; a condition of senility, mental illness, alcoholism or drug abuse; certain degenerative diseases that may be so advanced at the time of implantation that the expected useful life of the device is substantially decreased, and medical conditions that may affect postoperative management, such as alzheimer¿s disease and emphysema.The patient should be informed of the potential adverse effects (risks/complications) contained in this insert(see safety results / aes).Information on the proper implant site preparation, implant size selection, and the use of surgical instrumentation for the simplify® cervical artificial disc is provided in the surgical technique guide and should be reviewed prior to surgery.Preoperative planning may be used to estimate the required implant size and to assure that the appropriate range of sizes is available for surgery.Correct selection of the appropriate implant sizes is extremely important to assure the placement and function of the disc." "intraoperative: use aseptic technique when removing simplify®cervical artificial disc from the innermost packaging.Carefully inspect each device and its packaging for any signs of damage, including damage to the sterile barrier.Do not use the simplify® cervical artificial disc if the packaging is damaged or if the implant shows signs of damage.Ensure simplify® cervical artificial disc does not come into contact with hard objects that may damage the implant and render the implant functionally unreliable.Visual inspection of the prosthesis is recommended prior to implanting the device.If any part of the device appears damaged or not fully assembled, do not use.Take care not to over-distract the disc space.Excessive removal of endplate cortical bone may result in sub-optimal outcomes." "postoperative: patients should be instructed in postoperative care procedures and should be advised of the importance of adhering to these procedures for successful treatment with the device.Heavy lifting (greater than 20lbs) should be avoided for 6 weeks, and impact sports should be avoided for 3 months." "potential adverse effects of device on health below is a list of the potential adverse effects (e.G., complications) identified from the simplify® cervical artificial disc clinical study results, approved device labeling for other cervical total disc replacement devices, and published scientific literature including: (1) those associated with any general surgical procedure; (2) those associated with anterior cervical spine surgery; and (3) those associated with a cervical artificial disc device, including the simplify® cervical artificial disc.In addition to the risks listed below, there is also the risk that surgery may not be effective in relieving symptoms or may cause worsening of symptoms.Additional surgery may be required to correct some of the adverse effects." "general surgery risks general surgical risks are, but may not be limited to: infection/abscess/cyst, localized or systemic." "anterior cervical surgery risks anterior cervical surgical risks are, but may not be limited to: infection/abscess/cyst, localized or systemic." "cervical artificial disc risks specific to cervical artificial discs, including the simplify® cervical artificial disc, are but may not be limited to: infection/abscess/cyst, localized or systemic.Allergic reaction to the implant materials.Implant failure, device instability, improper device sizing, wear debris, removal, revision, reoperation or supplemental fixation of the disc.Osteolysis, bone loss, or bone resorption." 32226-e-2023-03.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLIFY CERVICAL ARTIFICIAL DISC
Common Device Name
PROSTHESIS INTERVERTEBRAL DISC
Manufacturer (Section D)
NUVASIVE INC.
7475 lusk blvd.
san diego CA 92121
Manufacturer (Section G)
NUVASIVE INC.
7475 lusk blvd.
san diego CA 92121
Manufacturer Contact
whitney ginter
7475 lusk blvd.
san diego, CA 92121
MDR Report Key20571556
Report Number2031966-2024-00290
Device Sequence Number20971307
Product Code MJO
UDI-Device Identifier00843285100078
UDI-Public843285100078
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P200022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberMD-5L
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2024
Initial Report FDA Received Date10/30/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age39 YR
Patient SexMale
-
-