MAUDE Adverse Event Report: MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PLUS PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PLUS PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Catalog Number 3501440
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 10/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient who underwent breast augmentation with mentor smooth round moderate plus profile 275cc saline protheses experienced bilateral deflation and right sided capsular contracture post procedure.The left implant had been previously replaced for a deflation in a previous event.The right implant had notably hardened prior to deflation.As a result, implant replacement was performed on (b)(6) 2024.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation / capsular contracture d4: udi: as the serial number for the device involved in the event was not provided, the full udi is currently not available.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
The mentor failure analysis lab received the device for evaluation on november 29, 2024.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
The investigation of the returned device was completed by the failure analysis lab on december 26, 2024.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection, leak testing, and microscopic examination of the returned device.Visual analysis of the returned sample revealed that the breast implant had a crease/fold extended from the anterior to the posterior view.Leak testing was performed, according to the mentor procedures, and it identified a tear within the crease/fold on the anterior view measuring approximately 0.2 cm.A microscopic examination was performed, and instrument damage was not found at the edges of the tear.The evaluation determined that the possible cause of the rupture was consistent with normal wear.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in deflation at a later time.Breast implants may also simply wear out over time.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SMOOTH ROUND MODERATE PLUS PROFILE
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key20571847
Report Number1645337-2024-12782
Device Sequence Number8096552
Product Code FWM
UDI-Device Identifier00081317001980
UDI-Public00081317001980
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2022
Device Catalogue Number3501440
Device Lot Number7589404
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/11/2024
Supplement Date Received by Manufacturer11/29/2024
12/26/2024
Initial Report FDA Received Date10/30/2024
Supplement Report FDA Received Date12/19/2024
12/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/30/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-