The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd module with drive pcb.In addition, our technician confirmed that the light guide cable coating damage, the insertion flexible tube compressed (short in length), the remote control buttons cracked, the objective lens cracked, the segment corroded, the bending rubber leak, the brand plate worn out, the operation channel (primary) resistance, and the insertion flexible tube discolored; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|