| Model Number |
SY60WF |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported with a description of the haptic was stuck under the injector¿s plunger during delivery of the intraocular lens (iol).There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned for analysis.The associated company lens with 25.5 diopter is not qualified for use with the company cartridge.The qualified diopter range for the company cartridge is 6.0 to 25.0.The reported complaint was not observed as no sample was returned for analysis; however, based on the information provided by the customer, the most likely root cause is failure to follow instructions for use (ifu), as the surgeon states the use of non-qualified cartridge.The use of an unqualified combination may cause damage to the intraocular lens (iol) and potential complications during the implantation process.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|
| |
|