• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MSD DEGGENDORF MFG DIVERGENCE¿ ANTERIOR CERVICAL FUSION SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED F

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MSD DEGGENDORF MFG DIVERGENCE¿ ANTERIOR CERVICAL FUSION SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED F Back to Search Results
Model Number G6626526
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/02/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Radiographic image review states, "sagittal computed tomography reconstruction shows a screw in the anterior prevertebral space not engaged in the bone.Sagittal computed tomography view shows superior screw of standalone interbody connected device partially engaged in bone; whole graft appears positioned anteriorly." medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient who underwent c5/6 ant erior cervical discectomy and fusion for cervical spondylotic myelopathy.It was reported that the caudal screw was cut out after surgery.Additional surgery planned to remove the screw.There were no patient symptoms reported.There were no further complications reported regarding the event.Additional information received that the cage has moved from the right place to wrong place.Revision surgery was performed on (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIVERGENCE¿ ANTERIOR CERVICAL FUSION SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED F
Manufacturer (Section D)
MSD DEGGENDORF MFG
wertstrasse 17
deggendorf 94469
GM  94469
Manufacturer (Section G)
MSD DEGGENDORF MFG
wertstrasse 17
deggendorf 94469
GM   94469
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key20576303
Report Number1030489-2024-01403
Device Sequence Number18185215
Product Code OVE
UDI-Device Identifier00643169337497
UDI-Public00643169337497
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K141599
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG6626526
Device Catalogue NumberG6626526
Device Lot NumberH5926450
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2024
Initial Report FDA Received Date10/31/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-