| Catalog Number |
A041122110112 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
10/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device was not returned for evaluation.However, the lot number was provided and retained-product testing and/or dhr review are planned.The results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that a protaper fin.21mm/f1 *not ce* file broke during use.The broken part could not be removed.Further treatment is an option, it was suggested to the patient to see a specialist to remove the broken part and undergo the root canal treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
Involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.Provided batch number has no corresponding in our system.Right batch number is unknown, dhr cannot be reviewed.Root causes are not identified.We will track this kind of event and monitor the trend.Potential root causes may be incorrect technique (motor setting not communicated), overuse (number of uses not communicated), excessive wear, or material issue (no analysis of the broken fragments possible).
|
| |
|