MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3501501BC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a 150cc mentor memorygel breast implant prosthesis ruptured during a breast reconstruction surgery as it was being implanted into the patient.However, the surgery commenced without any reported delays and a new implant was used to complete the implantation.No adverse events or patient consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.D4: udi: product made prior to udi compliance date.Udi not required.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: n/a.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On november 19, 2024, the mentor analysis lab received the suspect medical device for evaluation.On november 28, 2024, mentor completed an evaluation on the returned device.Mentor then conducted a visual inspection and microscopic examination of the device.Visual analysis of the returned sample determined that the breast implant was found to have a tear on the anterior view measuring approximately 4.8 cm.Microscopic examination was performed on the edges of the rupture, and parallel striations were found in an area of the tear, measuring approximately 0.2 cm.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.The cause of the rupture in the remaining area of the tear could not be identified.Manufacturer¿s reference number: (b)(4).
|
| |
|