| Catalog Number |
MC1005T |
| Medical Device Problem Codes |
Failure to Osseointegrate (1863); Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Codes |
Pain (1994); Metal Related Pathology (4530)
|
| Date of Event |
10/06/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a revision surgery was performed 1-2 years post-operatively to address metallosis and non-fusion of an roi-c patient.The patient was experiencing pain prior to the revision.Using spect-ct, the surgeon determined that the cage was loose and that there was a visible amount of metallosis.This is report one of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: the remainder of the udi number is unknown because the lot number is not available.Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information in h6: investigation type, findings, and conclusions.Device evaluation: the device was not returned and no photos or x-rays were provided, so an evaluation is unable to be performed.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to inadequate vertebral body end plate preparation.This is assuming that the anchoring plates were correctly installed, and without knowing the patient¿s smoker status or having a diabetic condition.If all residual disc material was not removed from the disc space, especially around the area of the hole in the spacer in which the bone replacement matrix is added, then bone will be unable to grow from one endplate through the center of the spacer and to the opposite endplate.Dhr review: the lot number was not reported and the device was not returned, therefore a dhr review is unable to be performed.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.Reference report 3004788213-2024-00083.
|
| |
|
Event or Problem Description
|
|
It was reported that a revision surgery was performed 1-2 years post-operatively to address metallosis and non-fusion of an roi-c patient.The patient was experiencing pain prior to the revision.Using spect-ct, the surgeon determined that the cage was loose and that there was a visible amount of metallosis.This is report one of two for this event.
|
| |
|